  • Stereotactic Neuronavigation Based Only on Three-Dimensional Rotational Angiography Using Surface-Based Facial Registration

    Final Number:
    280

    Authors:
    David A. Stidd MD MS; Joshua Thomas Wewel MD; Ali J. Ghods BS; Stephan Munich BS, MD; Kiffon M. Keigher N.P.; Heike Theessen; Roham Moftakhar MD; Demetrius K. Lopes MD

    Study Design:
    Laboratory Investigation

    Subject Category:
    Cerebrovascular

    Meeting: AANS/CNS Cerebrovascular Section 2014 Annual Meeting

    Introduction: Cerebrovascular lesions can have complicated abnormal anatomy not completely characterized by CT or MR angiography. 3D rotational angiography provides superior spatial and temporal resolution, but catheter angiograms are not easily registered to the patient, limiting the use of these images as a source for neuronavigation. Digital subtraction angiography (DSA) contains not only vascular anatomy, but also facial surface anatomy data. We report a novel technique to register DSA images using only the surface anatomy contained within the data set without fusing the DSA to other imaging modalities or the use of fiducials.

    Methods: A DSA scan was obtained of a fixed cadaver head with acquisitions of mask and contrast runs. The right ICA was injected prior to the contrast run with a 45% contrast solution diluted with water-soluble red liquid latex. One week later time, the head was registered to a Stealth navigation system loaded with the DSA images acquired earlier using facial surface anatomy (Figure 1). A right pterional craniotomy was performed and 10 different vascular landmarks were identified and measured for accuracy using the Stealth system. Neuronavigation based only on DSA was also used to treat a patient that presented with a distal MCA aneurysm.

    Results: The accuracy of the measurements for the cadaver model was 0.71 ± 0.25 mm, which is consistent to the 1.6 – 3.0 mm reported for neuronavigation. The DSA-based navigation assisted surgery for the distal MCA aneurysm aided in aneurysm localization, resulting in a small craniotomy and minimal brain dissection.

    Conclusions: This is the first example of frameless neuronavigation based on 3D catheter angiography registered by only surface anatomy data contained within the 3D DSA image set. This is an easily applied technique that is beneficial for accurately locating and reducing the dissection burden of vascular lesions.

    Patient Care: This new method will assist open cerebral vascular surgery.

    Learning Objectives: The reader will learn about a new method to register DSA images to a patient in the operating theater.