Skip to main content
  • Stereotactic Frame Registration with Intraoperative CT, Comparison with Standard CT

    Final Number:
    253

    Authors:
    Joshua M. Rosenow MD FACS; Andreas Haukas; Chad Jacobs; Mark J. Nolt PhD

    Study Design:
    Other

    Subject Category:
    Image Guided Applications

    Meeting: 2016 ASSFN Biennial Meeting

    Introduction:

    Stereotactic frame registration is often carried out using a standard CT scan obtained in the radiology department. With the availability of better intraoperative CT scanners the opportunity for streamlining the workflow for these cases now exists. We sought to determine if the AIRO scanner (Brainlab) was capable of performing scans adequate for frame registration and how the target coordinates provided by this registration compared with those obtained from a standard CT scanner

    Methods:

    Ten consecutive patients undergoing DBS surgery underwent standard CT scans to register the Leksell frame for DBS surgery. They also underwent low dose scans with the AIRO scanner. These scans were uploaded to Brainlab iPlan Stereotaxy 3.0 software. The frames were registered from these scans and the scans were merged with preoperative MRI. Commissural coordinates were obtained, as well as frame settings (x, y, z, arc and ring angles).

    Results:

    The frame was successfully registered in all cases, although more adjustment of window and level were required for the AIRO scans. All coordinates compared favorably. No pair of coordinates differed by more than 0.84 between scan registrations. The mean difference across all coordinates was 0.18. The largest mean difference between coordinate pairs was 0.21 for both the target Y and Z coordinates. There was no significant difference in the variation between scans for the AC, PC, or target.

    Conclusions:

    The AIRO scanner can capably register the Leksell frame and provide coordinates comparable to those obtained with a standard CT scan. Abandoning the trip to radiology may save time and effort as part of stereotactic case workflow. This may also reduce total patient radiation exposure by using the AIRO low dose sinus scan mode.

    Patient Care:

    This research impacts the patient flow and efficiency of cases involving stereotactic frames

    Learning Objectives:

    By the conclusion of this session, participants should be able to: 1. understand the use of intraoperative CT for frame registration 2. understand the impact of intraoperative CT on stereotactic case workflow