Skip to main content
  • A Biomechanical Evaluation of a Versatile and Novel Anterior Cervical Fusion Device Possessing Modular and Integrated Fixation Capabilities

    Final Number:
    284

    Authors:
    Ripul R. Panchal DO, FACS; John E. Wanebo MD; Anup Gandhi PhD; Chris Ferry MS; Sam Farmer; Kaushik Das MD

    Study Design:
    Laboratory Investigation

    Subject Category:

    Meeting: Section on Disorders of the Spine and Peripheral Nerves 2016 Annual Meeting

    Introduction: A novel ACDF (anterior cervical discectomy and fusion) construct possessing integrated screw and modular plate fixation (MPF) capabilities has been introduced in an effort to provide intraoperative versatility when selecting a stabilization mechanism (Fig 1). Such versatility allows the surgeon to switch between zero-profile, half-plate, and full-plate fixation without needing to substitute or reposition the cage in-situ. The objective of this study was to assess segmental rigidity achieved with the novel ACDF construct in comparison to traditional ACDF and anteroposterior cervical fusion constructs.

    Methods: Twelve human cervical spine specimens (C3-T1) were tested. Osseous integrity was confirmed via DEXA scans and radiographs. Specimens were then divided into three groups such that the mean bone quality across each group was consistent. The C3 and T1 vertebral bodies were potted. Each spine was first tested in an intact state. An anterior discectomy (C5/C6) was then performed, followed by sequential iterative construct instrumentation and testing (see Results for sequence). The three group protocol was executed such that each specimen received only zero-profile, half-plate, or full-plate fixation respectively (Alta ACDF System – Zimmer Biomet Spine). A 2Nm moment was applied in flexion/extension (FE), lateral bending (LB), and axial rotation (AR) using a six degree-of-freedom kinematics system. Segmental range-of-motion (ROM) was tracked using a motion analysis system. Mean ROM relative to intact conditions was measured.

    Results: See Table 1 and Figure 2.

    Conclusions: The full-plate and half-plate constructs both appeared advantageous in comparison to the traditional ACDF construct when used as a stand-alone device, demonstrating an inherent benefit of the MPF technology. Motion reductions with zero-profile fixation were not as robust; however, it can be argued that clinically significant stability was still achieved. Lastly, supplemental fixation with LMS appeared to be a clear leveling factor across all constructs, facilitating significant motion reduction in all principle directions.

    Patient Care: This research indicates that the discussed novel ACDF device, which can reduce intraoperative demand on the surgeon, increases segmental rigidity in comparison to traditional technologies, effectively supporting the improvement of surgical outcomes as well as long-term healing/fusion outcomes in ACDF.

    Learning Objectives: By the conclusion of this session, participants should be able to discuss/identify… 1) Benefits of in-situ fixation adjustment capabilities 2) Stabilization characteristics of a novel and versatile ACDF device 3) Patient demographics/pathologies for which the novel ACDF device could be advantageous 4) Similarities and differences between outcomes achieved with the novel ACDF device and traditional ACDF

    References:

We use cookies to improve the performance of our site, to analyze the traffic to our site, and to personalize your experience of the site. You can control cookies through your browser settings. Please find more information on the cookies used on our site. Privacy Policy