Skip to main content
  • RETROSPECTIVE STUDY OF USE OF ENTERPRISE® STENT IN THE TREATMENT OF RUPTURED WIDE-NECKED INTRACRANIAL ANEURYSMS IN NEUROENDOVASCULAR CENTER OF UNIVERSITY OF PUERTO RICO

    Final Number:
    134

    Authors:
    Hector T Rondon MD; Rafael Rodriguez-Mercado MD, FACS; Caleb Feliciano-Vals MD; Joel Enrique Pellot-Cestero

    Study Design:
    Clinical trial

    Subject Category:
    Aneurysm/Subarachnoid Hemorrhage

    Meeting: AANS/CNS Cerebrovascular Section 2018 Annual Meeting

    Introduction: Over the past decade, endovascular approach has become a safe alternative for the treatment of cerebral aneurysms, opening the possibility for treatment of different aneurysm variants, such as: wide neck, fusiform, and aneurysms in the distal segments of main arteries. Nevertheless, stent embolization of ruptured wide neck remains a challenge, since most of these patients have been referred to open surgery in the past.

    Methods: Retrospective review of a registry of 30 patients with ruptured cerebral aneurysms treated with stent assisted embolization and prospective review of their evolution from June 2008 to June 2017. We analyzed the anatomical features, localization, and complications years after treatment, including percentage of occlusion and recanalization.

    Results: In 70% of the cases were female; median age was 53 years, most frequent location was the carotid ophthalmic segment (40%), followed by posterior communicating artery (20%) and posterior circulation (15.5%). Average dome size was 7.8 mm and neck size was 5.10 mm. Following the Raymond classification at one year follow up, 90% had complete obliteration, 7% required repositioning, 1 patient had migration of the stent, 3 patients required re-embolization due to residual aneurysm, 2 presented bleeding during the procedure, 2 had hematoma at site of arterial puncture and 1 patient died.

    Conclusions: The use of Enterprise® stent for assistance in coil embolization of challenging ruptured wide neck aneurysms is effective and safe. Factors that add complexity to the treatment include the following: arteries with a sharp angle, arteries with a larger diameter at aneurysm area, vasospasm close to the aneurysm, tortuosity and hydrocephalus development post bleeding. High resolution fluoroscopy unit is needed for adequate visualization of the stent under fluoroscopy guidance. We do not recommend stent deployment in sharp angle zones or in large diameter in order to decrease the risk of stent migration.

    Patient Care: BE ABLE TO MAKE THE BEST DECISION IN THE CHOICE OF THE DEVICE TO BE USED IN CASES OF RUPTURED WIDE-NECKED INTRACRANIAL ANEURYSMS

    Learning Objectives: Evaluate the feasibility, safety and efficacy of the Enterprise stent in the treatment of 30 patients with ruptured aneurysms from 2008 to 2017