Skip to main content
  • Extraction of Broken Sacral Pedicle Screws Using a Left Handed Core Drilling Bit: A Technical Report

    Final Number:

    Authors:
    Suraj Sulhan; Kristopher Allen Lyon MD; David Garrett MD

    Study Design:
    Other

    Subject Category:

    Meeting: Congress of Neurological Surgeons 2018 Annual Meeting

    Introduction: Broken pedicle screws are a well-recognized complication of spinal instrumentation procedures, reportedly occurring in 0.5% to 11.2% of cases, and thought to be the result of excessive fatigue placed on the placed hardware. The resulting loss of fixation at these levels have been associated with pseudoarthrosis, instability, pain and poor surgical outcomes. These screws may also fracture within the pedicle itself, leaving broken screw threads in the pedicle wall and the decision on whether to remove the instrumentation is not fully agreed upon. Current methods include drilling a pilot hole in the fragment and using a screw extractor or by drilling a slot in the visible screw end to allow for the use of a normal screw driver. However, existing methods often involve further bone resection, weaken the involved pedicle or risk damage to the adjacent dura and neural tissue.

    Methods: A 40-year-old male who previously underwent an L5 to S1 laminectomy and fusion presented to our institution with worsening symptoms and CT evaluation yielded bilateral S1 pedicle screw fractures and non-union of fusion at the L5 to S1 levels. The patient underwent removal of the existing lumbosacral instrumentation and had remaining screw fragments within the pedicle. A left-handed core drilling screw extraction drill bit was then used to remove to remaining fragments.

    Results: Literature review of broken pedicle screws, hardware failure and screw extraction techniques was done. To our knowledge, there have been no cases reported of lumbosacral pedicle screw fragment extraction using our technique.

    Conclusions: Extraction of broken pedicle screw fragments, especially when threaded fragments remain within the pedicle, remain a technical challenge in Neurosurgery. The use of a left-handed core drilling screw extraction bit may facilitate safer extraction of hardware while keeping nearby bony and neural structures intact.

    Patient Care: Our aim is to provide use of another available set of instrumentation to offer a safer and less complex technique for screw fragment removal that may lead to less complications seen in existing techniques.

    Learning Objectives: By the conclusion of this session, participants should be able to: 1) Describe the occurrence rate and mechanism of broken lumbosacral pedicle screws. 2) Discuss current screw extraction techniques and possible associated complications that may occur. 3) Identify an effective and safer alternative to extraction of broken pedicle screw fragments and describe a case in which it’s used.

    References: Duncan, J.D. and J.D. MacDonald, Extraction of broken pedicle screws: technical note. Neurosurgery, 1998. 42(6): p. 1399-400. Elgafy, H., et al., Proposed alternative revision strategy for broken S1 pedicle screw: radiological study, review of the literature, and case reports. Spine J, 2013. 13(7): p. 796-802. Sewell, M.D., et al., Broken pedicle screw thread extraction from within the pedicle. Ann R Coll Surg Engl, 2016. 98(4): p. 285-6.

We use cookies to improve the performance of our site, to analyze the traffic to our site, and to personalize your experience of the site. You can control cookies through your browser settings. Please find more information on the cookies used on our site. Privacy Policy