Skip to main content
  • The Predictiveness of the Thermal Damage Estimate in Magnetic Resonance Guided Laser Induced Thermal Therapy

    Final Number:
    648

    Authors:
    Kiersten Frenchu; Nitesh V. Patel MD; Shabbar F. Danish MD

    Study Design:
    Other

    Subject Category:

    Meeting: Congress of Neurological Surgeons 2016 Annual Meeting

    Introduction: Magnetic Resonance Guided Laser Induced Thermal Therapy (MRgLITT) is a minimally invasive method to treat a wide range of intracranial pathologies. Ablation extent is estimated through the Arrhenius model and pixel-shift of ablated tissue on MRI. A thermal damage estimate (TDE) is plotted over a 2D cross section of the lesion containing the laser catheter. This study evaluates the accuracy and reliability of the TDE in predicting ablation extent.

    Methods: Medtronic Visualase system was used. Only cases with clear ablation borders and primarily single laser catheters were included. On the post-operative axial T1 contrast enhanced image, the slice corresponding to the intraoperative image for the TDE was acquired. The intraoperative TDE image was obtained from the Visualase console. OsiriX DICOM Viewer was utilized to for cross sectional ablated area on the post-op MRI. ImageJ was utilized to measure estimated area of ablation on the intraoperative TDE image. Two raters performed all measures. Statistical testing included Pearson correlation and the student’s t-test.

    Results: A total of 32 patients were included with an average age of 49. 62.5% were female. Pathology included primary and metastatic brain lesions, along with mesial temporal lobe sclerosis and chronic pain (cingulotomy). Measured ablated area strongly correlated between raters (r=0.93, p<0.0001). The measured TDE on intraoperative imaging also shared a strong correlation between the two raters (r=0.97, p<0.000001). There was no significant difference between TDEs and post-ablation ablated area (p=0.40) and TDE correlated to ablated area closely (r=0.84, p<0.001).

    Conclusions: The TDE is an accurate and reliable measure of actual ablated area in MRgLITT. Future studies should incorporate larger numbers of patients and assess the reliability of the TDE in cases where multiple lasers are used on the same lesion.

    Patient Care: This research helps to validate the use of the TDE as a measurement of ablation damage. It will allow doctors to be confident in the accuracy of this method of measurement and help reassure patients.

    Learning Objectives: The main learning objective is to determine whether the TDE is an accurate and reliable method of measuring laser ablation damage. This is the first known study examining the accuracy of the TDE using human intracranial cases.

    References: None