Skip to main content
  • Mechanomyography (MMG) for Intraoperative Assessment of Screw Placement During Instrumented Spine Surgery

    Final Number:
    1179

    Authors:
    Richard Rammo MD; Hesham Mostafa Zakaria MD; Kevin Reinard MD; Lonni Schultz PhD; Todd Aho; Muwaffak Abdulhak MD

    Study Design:
    Clinical Trial

    Subject Category:

    Meeting: Congress of Neurological Surgeons 2016 Annual Meeting

    Introduction: EMG is widely used to assess pedicle screw placement during spine surgery. Recent studies have questioned the reliability of EMG due to its low specificity.1-4 Mechanomyography (MMG) monitors the mechanical activity of muscles and is the mechanical counterpart of EMG. Unlike EMG, MMG is not susceptible to electrical interference or stimulus artifact. This study reports the effectiveness of MMG to intra-operatively assess cortical breach prior to screw insertion during lumbar spine instrumentation.

    Methods: Patients were enrolled in a prospective study comparing intraoperative MMG values with post-operative CT scan for the assessment of cortical breach of pedicle screws. Intraoperative MMG values were obtained after screw trajectories were piloted, but prior to screw insertion, by stimulating all four pedicle walls (medial, lateral, inferior, superior) with a 15mA stimulus. If a positive MMG response was observed, stimulation current was decreased until the response disappeared. The minimum level of current capable of eliciting an MMG response for all four walls of each screw hole was noted. CT scans were taken post-operatively with interpretation performed by a single radiologist.

    Results: 122 consecutive patients with 896 lumbar pedicle screw pilot holes were tested intra-operatively with MMG. 7.4% of screw trajectories were repositioned intra-operatively prior to screw insertion. Post-operative CT showed a breach rate of 14.94% (95%CI=11.97-17.92), with 2.25% (95%CI=1.02-3.47) being medial or inferior. All breaches were asymptomatic, and no screw revisions were required. Using a cutoff of >7mA, the breach rate for medial and inferior for pilot hole was 1.09% (95%CI=0.17-2.01), compared to 2.17 (95%CI=1.02-3.32) for the screw test.

    Conclusions: MMG cutoff values of >7mA decreases the overall breach rate and resulted in a 98.9% likelihood of no medial or inferior breach during lumbar instrumentation. All breaches experienced during the use of MMG were asymptomatic, and none revision surgery. Further studies are needed to directly compare EMG to MMG.

    Patient Care: It will improve patient care by decreasing the rate of pedicle breach in instrumented spine surgery, decreasing the rate of nerve injury and improving fixation.

    Learning Objectives: By the conclusion of this session, participants should be able to: 1) Describe the differences between electromyography and mechanomyography (MMG) for the detection of nerve injury during instrumentation 2) Discuss the use of MMG for prevention of pedicle breach

    References: 1. Lee CH, Kim HW, Kim HR, Lee CY, Kim JH, Sala F: Can triggered electromyography thresholds assure accurate pedicle screw placements? A systematic review and meta-analysis of diagnostic test accuracy. Clin Neurophysiol 126:2019-2025, 2015 2. Mikula AL, Williams SK, Anderson PA: The use of intraoperative triggered electromyography to detect misplaced pedicle screws: a systematic review and meta-analysis. J Neurosurg Spine:1-15, 2015 3. Parker SL, Amin AG, Farber SH, McGirt MJ, Sciubba DM, Wolinsky JP, et al: Ability of electromyographic monitoring to determine the presence of malpositioned pedicle screws in the lumbosacral spine: analysis of 2450 consecutively placed screws. J Neurosurg Spine 15:130-135, 2011 4. Wang MY, Pineiro G, Mummaneni PV: Stimulus-evoked electromyography testing of percutaneous pedicle screws for the detection of pedicle breaches: a clinical study of 409 screws in 93 patients. J Neurosurg Spine 13:600-605, 2010

We use cookies to improve the performance of our site, to analyze the traffic to our site, and to personalize your experience of the site. You can control cookies through your browser settings. Please find more information on the cookies used on our site. Privacy Policy