  • Transorbital Ultrasound Measurement as a Non-invasive Marker of Intracranial Pressure (ICP)

    Final Number:
    125

    Authors:
    Llewellyn Padayachy MD, PhD; Graham A G Fieggen

    Study Design:
    Clinical Trial

    Subject Category:

    Meeting: Congress of Neurological Surgeons 2016 Annual Meeting

    Introduction: Reliable assessment of intracranial pressure (ICP) remains crucial in managing neurosurgical conditions in children. The present study examined the relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children.

    Methods: This was a prospective, observational study comparing ONSD measurement to invasive ICP measurement. All patients were under general anaesthesia and physiological variables including systolic and diastolic blood pressure, mean arterial pressure, pulse rate, temperature, respiratory rate and end tidal carbon dioxide level were recorded at the time of ONSD measurement. The ONSD measurements were analysed for intra- and inter-observer variability. The diagnostic accuracy of ONSD measurement for detecting ICP at thresholds of 20, 15, 10 and 5 mmHg was analysed, including evaluation of age-related thresholds in children.

    Results: 196 children were evaluated. The median age was 36 months (IQR 8-82). Etiology included, hydrocephalus (52.9%), traumatic brain injury (17.2%), tumour (9.2%), craniosynostosis (9.2%), cystic malformation (6.9%), other (4.6%). ONSD measurement demonstrated good correlation with ICP across the entire patient cohort (r = 0.66, p < 0.001), but was better in children > 1 year or with a closed anterior fontanelle (AF) (r = 0.7, p < 0.001). The ONSD values with the best diagnostic accuracy for detecting ICP above 20 mmHg in older children was 5.75 mm, with a sensitivity of 85.9%, specificity of 70.4% , PPV of 77.5%, NPV of 80.9% and AUROC of 0.78. The diagnostic OR was 14.5. Diagnostic testing was performed at ICP thresholds of 15, 10 and 5 mmHg as well.

    Conclusions: Measurement of the ONSD is a sensitive surrogate marker of raised ICP with limited specificity. This relationship was more reliable in older children, particularly when the AF was closed. This study provides the first data supporting the relationship between ONSD measurement and invasively measured ICP at various thresholds.

    Patient Care: Provides reference ranges for a non-invasive technique of assessing ICP, which are invaluable in improving the applicability and diagnostic accuracy of this technique.

    Learning Objectives: evaluation of non-invasive ICP assessment